14. Servicing
Inspection measures
Cancel switch/WC
Inspection interval: 12 months
Is the housing undamaged and free of
contamination?
Can a WC call be cancelled using the grey
cancel button? Check both buttons.
Do the LEDs next to the buttons light up
when a WC call that can be cancelled with
this cancel switch is activated?
If the optional "Call tone" function is acti-
vated via configuration on this switch,
check additionally: Does the switch emit a
call tone after an overflow time during ac-
tivated staff presence if a call has been in-
itiated in another room of the ward?
Overflow time according to operating
mode and configuration in the SystemOr-
ganizer.
Cardiac alarm switch
Inspection interval: 12 months
Is the housing undamaged and free of
contamination?
Can an alarm be triggered by pressing the
blue alarm button if presence is activated
in the room? Check both buttons.
Do the location and reassurance lights of
the cardiac alarm switch work?
Does the switch emit an acoustic acknowl-
edgement of the call initiation?
78
Order no.
77 0213 00 A
77 0213 00 C
77 0213 00 F
77 0214 00 A
77 0214 00 C
77 0214 00 F
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
77 0213 00 A
77 0214 00 A