Tunstall FlamencoIP Technical Manual page 80

Hide thumbs Also See for FlamencoIP:
Table of Contents

Advertisement

14. Servicing
Inspection measures
Cancel switch/WC
Inspection interval: 12 months
 Is the housing undamaged and free of
contamination?
 Can a WC call be cancelled using the grey
cancel button? Check both buttons.
 Do the LEDs next to the buttons light up
when a WC call that can be cancelled with
this cancel switch is activated?
 If the optional "Call tone" function is acti-
vated via configuration on this switch,
check additionally: Does the switch emit a
call tone after an overflow time during ac-
tivated staff presence if a call has been in-
itiated in another room of the ward?
Overflow time according to operating
mode and configuration in the SystemOr-
ganizer.
Cardiac alarm switch
Inspection interval: 12 months
 Is the housing undamaged and free of
contamination?
 Can an alarm be triggered by pressing the
blue alarm button if presence is activated
in the room? Check both buttons.
 Do the location and reassurance lights of
the cardiac alarm switch work?
 Does the switch emit an acoustic acknowl-
edgement of the call initiation?
78
Order no.
77 0213 00 A
77 0213 00 C
77 0213 00 F
77 0214 00 A
77 0214 00 C
77 0214 00 F
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)