14. Servicing
Functions
CT
ComStation
Flamenco
Inspection interval: 3 months
 Are the housing and connection cable un-
damaged and free of contamination?
 Is the plug of the connection cable firmly
seated in the socket and secured with
both screws?
 Is the display easily readable?
 Is a fault in the nurse call system shown
on the display? If so, locate the fault and
remedy it.
 Can presence 1 be activated and deacti-
vated using the green presence key?
Does the reminder light work?
 Can presence 2 be activated and deacti-
vated using the yellow presence key?
Does the reminder light work?
 If presence is activated: Can a cardiac
alarm be triggered by pressing the blue
alarm key? Prerequisite: According to the
configuration, the alarm key is available
on the ComStation
 Do the location (in case of presence) and
reassurance lights of the blue alarm key
work?
 Can a call be triggered by pressing the red
call switch?
 Do the location and reassurance lights of
the red call key work?
 If presence is activated: Does the Com-
CT
Station
emit a call tone after an overflow
time if a call has been trigged in another
room of the ward? Overflow time accord-
ing to operating mode and configuration in
the SystemOrganizer.
 Are the call type and call location shown
on the display?
Continued on the next page.
98
CT
.
Order no.
77 0606 00
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)