14. Servicing
Functions
Self-releasing adapter patient unit
Inspection interval: 3 months
 Are the plug and cable undamaged and
free of contamination?
 Does the patient handset work when the
self-relasing adapter is connected?
Self-releasing adapter pear push
switch
Like 74 0812 50, but for pear push switch.
Large-surface pneumatic switch
plug-in connection to connection socket
Inspection interval: 3 months
 Are the housing, button, connection cable
and plug undamaged and free of contam-
ination?
 Can a call be triggered by pressing the red
trigger button?
 Do the location and reassurance lights
work?
Sound detector
plug-in connection to connection socket
Inspection interval: 3 months
 Are the housing, keys, controllers, con-
nection cable and plug undamaged and
free of contamination?
 Function test according to the provided
documentation of the manufacturer.
Connection cable for call devices
plug-in connection to connection socket
Inspection interval: 3 months
 Are the cable and plug undamaged and
free of contamination?
 Does the connected call device trigger a
call within the nurse call system?
 Function test of the connected call device
according to the documentation of the call
device manufacturer.