14. Servicing
14.8
Radio call devices
Functions
Radio receiver-T
plug-in connection to connection socket
Recommendation of the manufacturer:
Weekly inspection with test call reception
under observance of the display elements
and call forwarding. If radio reception interfer-
ence is suspected, the radio range must be
checked as during initial start-up.
Are the housing, connection cable and
plug undamaged and free of contamina-
tion?
Does the LED indicator show a fault? If so,
remedy the fault. Example: If the LED on
the radio receiver flashes red, the battery
of an assigned radio transmitter is low.
Make sure the battery is changed.
Function test according to the provided
documentation of the manufacturer
MyAmie
to be assigned to radio receiver-T
Recommendation of the manufacturer of the
radio receiver-T: Weekly inspection with test
call reception under observance of the dis-
play elements and call forwarding.
Are the housing and button undamaged
and free of contamination?
Replace the neck cord or wrist strap ac-
cording to the hygiene regulations of the
hospital.
Can a call be triggered by pressing the call
button?
Does the red LED light up after the call
button is pressed? If it flashes, the battery
is low. Replace MyAmie.
Check whether MyAmie works across the
entire intended area of application (range
test).
92
Order no.
Z 00 8202 33
P68007/02
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)