Tunstall FlamencoIP Technical Manual page 95

Hide thumbs Also See for FlamencoIP:
Table of Contents

Advertisement

Functions
iVi
to be assigned to radio receiver-T
Recommendation of the manufacturer of the
radio receiver-T: Weekly inspection with test
call reception under observance of the dis-
play elements and call forwarding.
 Are the housing and buttons undamaged
and free of contamination?
 Replace the neck cord according to the
hygiene regulations of the hospital.
 If the LED flashes orange, a fault is pres-
ent. Remedy the fault.
 Can a call be triggered by pressing the call
button?
 Function test of the fall trigger according
to the user instructions of the iVi.
 Does the LED light up red after the call
switch is pressed? If it flashes red, the bat-
tery is low. Replace the battery. Service
life of the battery: approx. 12 months
 Check whether the iVi works across the
entire intended area of application (range
test).
Universal sensor
to be assigned to radio receiver-T
Recommendation of the manufacturer of the
radio receiver T: Weekly inspection with test
call reception under observance of the dis-
play elements and call forwarding.
 Are the housing and connection cable un-
damaged and free of contamination?
 Can a call be triggered by a triggering of
the connected call device?
 Function test according to the documenta-
tion for the connected call device.
 If the radio receiver of the universal sen-
sor indicates a low battery, the battery
must be replaced.
 Check whether the universal sensor
works across the entire intended area of
application (range test).
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
Order no.
P68007/47