Troubleshooting Flowcharts; Baseline Noise - Waters Prep Series Installation And Maintenance Manual

Table of Contents

Advertisement

General PrepLC troubleshooting (Continued)
Symptom
System controller
screen is blank
Increased system
backpressure during
injection using
Rheodyne injector

Troubleshooting flowcharts

The troubleshooting flowcharts in this section help isolate possible causes of
these problems:
High system pressure

Baseline noise

Erratic or incorrect retention times
Poor peak resolution
Column
If the troubleshooting flowcharts fail to address your problem, see
"Chromatography troubleshooting" on page
High system pressure
System pressure reference point
Choice of mobile phase, column, flow rate, and temperature can influence
system pressure, and it can vary significantly from one method to another.
You must therefore identify a pressure reference point. Each time you start a
new method, record the system pressure. Thereafter, to diagnose a problem
with system pressure, compare current system pressure to the reference
pressure you established.
Possible cause
Screen saver feature
active
Normal condition. When
the injector is in the
Inject position, the
sample loop generates
additional system
backpressure.
Corrective action
Press any function key to
display the screen.
To decrease system
backpressure, after
loading the sample,
return the injector to the
LOAD position.
3-16.
Troubleshooting
3-5