Gradient table values (Continued)
Time
Flow
45.00
1.00
To perform a solvent proportioning valve pair test:
1.
Go to the OPERATE METHODS screen.
2.
Press the START RUN screen key.
3.
When the detector baseline is stable at 0%, do the following:
a.
Change the detector sensitivity to 0.1 AUFS.
b.
Set the baseline to 10% full scale by adjusting the recorder's zero
adjustment.
c.
A 5% change on the recorder now equals a 1% change in
composition.
Evaluate results
The
figure "Test results with all valves working" on page 3-46
results obtained from this test when all valves are working.
Valve failure
A valve failure shows as a difference in height between the sections common
to that valve.
For example, if Valve A is malfunctioning, Peaks 1 and 2 (the only two
containing contributions from Valve A) are irregular when compared to Peaks
3 and 4. Similarly, if Valve D is bad, Peaks 2 and 4 are irregular.
%A
%B
0
100
%C
%D
0
0
is typical of the
Performance tests
Curve
11
3-45