High system pressure troubleshooting
High system
pressure
Set flow to 0.0
mL/min and
vent system
Pressure
reduced?
Yes
Remove column.
replace w/union.
run system.
Pressure
reduced?
No
Release
injector outlet
fitting
Pressure
reduced?
No
Release
injector inlet
fitting
Pressure
reduced?
No
Adjust
pressure
transducer
Yes
Troubleshoot
column
Yes
Troubleshoot
detector
Yes
Troubleshoot
injector
No
Pressure
reduced?
Yes
Call Waters
technical
service
Troubleshooting
Replace
pressure
transducer
3-7