B Specifications - Waters Prep Series Installation And Maintenance Manual

Table of Contents

Advertisement

B
Specifications
This appendix contains specifications for the following:
Operational
Electrical
Environmental
Operational specifications
Item
Number of solvents One to four
Modes of operation
Operating flow
range
Composition range
Compositional
accuracy
Automatic solvent
Sparging
Pressure maximum 2000 psi (140.6 kg/cm
Programmable
pressure limits
Specification
Gradient, isocratic, and flow programming
Programmable from 1 to 150 mL/min in
0.01-increments (500-µL heads) and from 4 to 300
mL/min in 0.01-increments (1000-µL heads).
0 to 100% programmable in 1%-increments for each of
four reservoirs: A, B, C, and D. Total composition
must sum to 100%.
Better than 3.0%.
Standard, helium gas, input gas pressure range 50 to
150 psi (3.5 to 10.5 Kg/cm
to 100 mL/min, programmable in 1-mL/min
increments.
(1000-µL heads); 4000 psi (281.2 Kg/cm
flow range (1 to 150 mL/min) with 500-µL pump
heads.
Upper: 51 to 2000 psi (3.6 to 281.2 kg/cm
Lower: 0 to 3950 (0 to 277.7 kg/cm
2
). Helium flow rate range 0
2
) over entire flow rate range
2
2
)
) over entire
2
)
B-1