Waters Prep Series Installation And Maintenance Manual page 73

Table of Contents

Advertisement

Column rack connections (Continued)
Column rack
Type of
connection
tubing
Purge outlet
Stainless
Fraction
PTFE
outlets
Waste
PTFE
The fluid handling unit connects to the column with a 0.040-inch ID tubing
(Startup Kit) installed between the pump outlet and the Inlet bulkhead
connector on the column rack's front panel.
Pump outlet connection to the column rack inlet (Delta-Prep system)
Column rack
Drip tray
To connect the pump and column rack:
1.
Cut a length of 0.040-inch stainless steel tubing (Startup Kit) that
reaches from the pump outlet to the Inlet connector on the column rack's
front panel.
Use excess tubing, forming it into several coils, as shown in the
Tip:
figure "Pump outlet connection to the column rack inlet (Delta-Prep
Tubing ID
Connect to
0.040
Waste container
0.062
Fraction containers
0.062
Waste container
Setting up the plunger wash
Controller
2487 Detector
Fluid handling unit
Drip tray
2-39