Pump troubleshooting (Continued)
Symptom
Erratic flow/pump
pulsations
Squeak
Injector troubleshooting
Before using the troubleshooting tables in this section, use those that appear
earlier in the chapter to identify the part of the system in which the problem
originates.
3-26
Troubleshooting, Testing, and Maintenance
Possible cause
Gas dissolved in mobile
phase
Air bubble in pump
head
Dirty check valve(s)
Faulty check valve(s)
Plugged solvent
reservoir filter
Kinked or bent tubing
Leaky plunger seals
Broken plunger
Defective solvent
proportioning valve
Pump seals not wetted
Binding plunger seal
Corrective action
Sparge solvents.
Prime pump to remove
bubble.
Clean check valve(s).
Replace check valve(s).
Clean or replace the
filter.
Remove bend from
tubing. If symptoms
continue, replace
tubing.
Replace seals.
Replace damaged
plunger.
Call WatersTechnical
Service.
Prime pump.
Replace plunger seal
assembly.