all screws are tight.
7.
With port 1 pointing upward, hold the valve for mounting.
8.
Place the valve in its slot, and fasten it.
9.
Reconnect all tubing to the valve.
10. Perform a needle wash by right-clicking in the Run Sample control panel
or via the Console Control menu.

Installing a sample loop

The instrument is fitted with a 100-µL sample loop as standard equipment.
You can install a different sample loop provided you select the proper
combination of syringe and tubing.
Connect the loop between ports 2 and 5 of the injection valve. When installing
a loop of nonstandard volume, from the Configure menu of the 2707 Console
click Volumes... and adapt settings.
Tip:
Calculate maximum injection volumes using the following
equations:
Full loop – Injection volume = loop volume
Partial loop needle overfill – Maximum injection volume = 50% of
loop volume
Partial loop – Maximum injection volume = (loop volume - [3 ×
needle volume])/2

Replacing the sample needle

To replace the sample needle
1.
From the Maintain menu in the 2707 Console, click Replace and the
needle.
Result:
2.
Follow the instructions in the Replace Needle warning box, and then
click OK.
3.
Loosen the needle connection nut.
3-6
Maintaining the Autosampler
Tighten each screw a half-turn each in rotation until
The needle moves to the exchange position.