Maintenance

All maintenance procedures described in this manual can be safely performed by
WA R N I N G
qualified personnel. Maintenance not covered in this manual should be performed
only by an Agilent Technologies representative.
Switch OFF the power and disconnect the power cord from the power supply before
performing any maintenance procedure that requires removal of any panel or cover,
or disassembly of any interior instrument component.
Removal of protective covers marked with the High Voltage warning symbol shown
at left can result in a safety hazard.

Cleaning the Well-plate Handler

BIOHAZARD Wear gloves during any cleaning procedure that could involve contact
WA R N I N G
with hazardous materials or fluids.
Periodically, you should clean the outside surfaces of the Well-plate Handler
using a cloth or sponge dampened with water, alcohol, glass cleaner, or a mild
soap solution diluted with water. If using glass cleaner or mild soap, wipe with
a damp cloth or sponge after cleaning to remove any residue. Do not use
abrasive cleaners. Do not spray cleaner onto the instrument. Clean the finger
pads of the gripper with alcohol or other residue-free solvent.
Should fluids spill on the base of the Well-plate Handler, the fluid is directed
away from the rotating base by means of a fluid barrier located beneath it.
Wipe up any spills immediately. Clean only the exterior of the unit. Never
remove any instrument panels to clean the inside of the instrument. Do not
allow excess water or other fluids to drip inside the instrument.
Agilent 1200 Series Sample Capacity Extension User Manual
Troubleshooting and Error Handling
5
95