Cleaning Electrodes after a Chip Run

Cleaning Electrodes after a Chip Run
When the assay is complete, immediately remove the used chip from the
Agilent 2100 Bioanalyzer and dispose of it according to good laboratory
practice. Then perform the following procedure to ensure that the electrodes
are clean (i.e. no residues are left over from the previous assay).
Use a new electrode cleaner with each new kit.
N O T E
Leak currents between electrodes
C AU T I O N
Liquid spill may cause leak currents between the electrodes.
1 Slowly fill one of the wells of the electrode cleaner with 350 μl deionized
2 Open the lid and place the electrode cleaner in the Agilent 2100
3 Close the lid and leave it closed for about 10 seconds.
4 Open the lid and remove the electrode cleaner.
5 Wait another 10 seconds to allow the water on the electrodes to evaporate
Replace the used electrode cleaner with each new kit.
N O T E
When switching between different assays, a more thorough cleaning may be required.
N O T E
For more details please refer to the "Maintenance and Troubleshooting Guide" which
is part of the Online Help of the 2100 Expert software.
20
Never fill too much water in the electrode cleaner.
analysis-grade water.
Bioanalyzer.
before closing the lid.