Accessing Detector Signal Boards; Replacing Detector Signal Boards - Agilent Technologies 7820A Service Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

Descripti
on
uECD
FID
NPD
TCD
FPD

Accessing detector signal boards

Replacing detector signal boards

Agilent 7820 GC Service
41
FPD
G3435-60010
Table
Detector signal electrical connectors
P1
P2
Communicati
Electromet
on buss
er
Communicati
Ignitor
on buss
Communicati
Bead
on buss
current
Communicati
on buss
Communicati
on buss
The front and back detector signal boards are located under the
right-hand side panel.
1 Remove the detector top cover and the electronics top cover.
Remove one screw on the side of the right hand side panel.
Loosen one screw and remove one screw on the back of the GC.
Slide the panel toward the back of the GC and lift off. Place the
tip of the driver in one of the top ventilation slots and push back.
1 Cool down the oven.
Turn off the GC main power switch and disconnect its power
cord.
F-DET or B-DET. Also uses AUX