Performing A First Time Calibration Of The Roll And Pitch Measurements Of A Px Multisensor - Simrad TV80 Reference Manual

Catch monitoring software
Table of Contents

Advertisement

Simrad TV80 Reference Manual
4
Observe the numerical presentation of the sensor depth, and allow the reading to
stabilize itself.
5
On the
menu, select
Setup
6
Select the input system from the available list.
7
Select
Start Calibration.
8
Wait for the system to do the calibration.
The dialog box will close when the calibration process is finished.
9
Observe the numerical presentation of the sensor depth, and verify that it reads 1 meter.
10 Select
to save the chosen settings and close the dialog box.
Close
11 Retrieve the sensor from the water.
Related topics
Calibration dialog box, page 214
Performing a first time calibration of the roll and pitch
measurements of a PX MultiSensor
If you wish to use the PX Multisensor to measure the pitch and roll of the trawl doors, you
must calibrate the sensor. You can calibrate a sensor in the
dialog box.
Prerequisites
Calibration must be done with the sensor mounted in its installation adapter on the trawl
door(s).
The
PX MultiSensor Configurator
• You have connected the sensor to you computer through the configurator cable.
• There is a receiver unit connected to the system.
Context
Whenever a sensor is mounted on a trawl door, the pitch and roll measurements need to be
calibrated. If you do not calibrate your sensors you will not know if the pitch and roll
readings are correct. If existing sensors have been replaced, or new sensors have been
installed, a new calibration is always required.
128
.
Calibration
dialog box is only available if:
PX MultiSensor Configurator