4. Loosen the four screws securing the LVPS cover.
5. Disconnect the fan harness.
6. Remove the fan and the mounting plate.
7. Pull up on the four rubber isolators to separate the fan from the mounting plate.
8. To reinstall the fan, perform the steps in reverse order. When reinstalling the fan, make sure
the air flow directional arrow is facing up.

Integrator assembly

The integrator rod glass should be flush with the output aperture with the black finish. If the glass is
damaged, contact Christie Technical Support staff immediately. Always wear gloves when you handle
the integrator assembly.
The integrator assembly is mounted in the light tube of the IOS. It runs through a firewall cutout
that separates the lamp housing from the light engine.
1. Turn the projector off and then disconnect it from AC power.
2. Remove the top lid. See
3. Remove the service door. See
4. Remove the six integrator housing screws.
5. Remove the housing.
6. Remove the two rotational integrator adjustment screws.
7. Slide the integrator rod back and tilt it to remove it from the casting.

Removing the contrast aperture

The contrast aperture is located after the fold mirror in the light path within the IOS.
1. Turn the projector off and then disconnect it from AC power.
2. Remove the top lid. See
CP2208 Service Manual
020-101917-01 Rev. 1 (11-2015)
Removing the top lid
Removing the service door
Removing the top lid
Parts and Module Replacement
on page 33.
on page 38.
on page 33.
60