Managing an ITEX Syringe and Trap Assemble

Installing the ITEX Syringe and Trap into the ITEX Tool
Thermo Scientific
This section provides information for installing the ITEX syringe and trap into the ITEX
Tool, for removing the HS syringe from the ITEX tool, for setting the flush gas pressure.
See the related topic:
"Installing the ITEX Syringe and Trap into the ITEX Tool"
"Removing the HS syringe from the ITEX Tool"
"Flush Gas Considerations"
To install an ITEX syringe and trap into the ITEX tool
1. Assemble the ITEX Syringe with trap/needle following the order shows in
Figure 58. ITEX Syringe and Trap Assembling
Fitting
a
b
c
d
a. Slide the seal provided with the ITEX trap on the bottom of the trap up to reach the
stopper.
b. Carefully insert the ITEX trap into the ITEX syringe as far as goes.
c. Slide the fitting on the ITEX trap up to reach the tip of the syringe, then secure the
assemble by screwing the manually fitting into the syringe without overtightening.
d. Internal view of the assemble where the ITEX trap is coupled with the ITEX syringe.
2. Remove the retaining nut from the tool, locate the O-ring and put it over the syringe glass
barrel up to the flange. This is a protection for the glass flange. See
on
page 63
ITEX Trap
Seal
2
Installation of the TriPlus RSH System
Managing an ITEX Syringe and Trap Assemble
on
page 61
on
page 63
ITEX Syringe
Figure
TriPlus RSH Hardware Manual
Figure
58.
59.
61