TriPlus RSH Defining Object Positions

Teaching Fast Wash Station Position

Teaching Fast Wash Station Position

92
TriPlus RSH Hardware Manual
Figure 85. Teach Points for Three Solvent Reservoir and Position 1 to 3
The teach position of a Fast Wash Station is the lunette at the front of the module.
See
Figure
86.
Figure 86. Teach Points for Fast Wash Station
To define the position of the fast wash station
1. Select the access Extended User Level.
2. From the Options pull-up menu, select the item Setup | Modules | Wash Stations | Fast
Wash.
3. If required by the application, attach the magnetic vial transport ring to the lower needle
guide.
4. Define the X-, Y-, and Z- positions common to all objects. The teach position is located
at the front of the upper drawer. The lower needle guide should be centered in the
Lunette.