LAN Communication Problems Related to Network Interface Card
(NIC)
Table 21. LAN Network Interface Card Troubleshooting Guideline 1 (Sheet 1 of 2)
Symptom or Error Message
No LAN communication.
Thermo Scientific
A network interface controller (also known as a network interface card, network adapter, LAN
adapter and by similar terms) is a computer hardware component that connects a computer to
a computer network. It is called here in short form NIC.
Table 21
provides troubleshooting guidelines regarding the LAN Network Interface Card.
Possible Cause
Network cable not plugged in.
Network cable plugged in into the
wrong NIC.
IP address setting in the TriPlus
RSH wrongly assigned.
IP address setting in the CDS
software wrongly assigned.
Subnet Mask wrongly assigned.
Default Gateway wrongly set.
IP address on PC NIC not valid.
Wrong LAN cable.

LAN Communication Problems Related to Network Interface Card (NIC)

Recommended Action
Check connections.
It is possible that more than one NIC is present on
the PC.
Select network connections settings on the computer,
plug in the cable, and check if the status changes
from unplugged to connected.
Check the IP address entered on the Handheld
Controller (or Virtual Handheld Controller.
Check the IP address entered on the CDS software.
Ask network administrator for details.
Ask network administrator for details.
The IP address has to be in the correct range such
that both the instrument and the PC are in the same
network and subnetwork.
Straight wired Cat. 5 LAN cable is used for most
setups. The only exception is if a direct connection
from PC to instrument is used and the NIC does not
have Auto-negotiation or Auto-sensing (the
possibility to detect the correct transmission
parameters and set them accordingly between NICs).
The network administrator may support and check:
MDI-X
Auto MDI(X)
Ports are used within the network.Check if straight
or crossed wired cable (vice-versa) solves the
problems, or work with a corresponding LAN
adapter (MDI) to test the communication.
TriPlus RSH Hardware Manual
6
Troubleshooting