Manually Removing A Tool From The Head - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

Manually Removing a Tool from the Head

Thermo Scientific
Table 23. LED Status
TriPlus RSH Status
Successful Booting Process
Power Off
System booting
Software OS Initialized
Application Software Initialized
System Ready
Observing the status of the control LED is also important when restoring the network settings
to factory default.See the section
When a syringe plunger or needle is bent during an automated procedure, it is likely that the
tool can no longer be parked in the Tool Station.
The de-coupling mechanisms can be blocked by the distorted part of the syringe.
In such situations it is important that you can manually remove the Tool from the Head.
WARNING Turn off the power to the TriPlus RSH before any repair steps are taken.
To remove a tool from the head of the TriPlus RSH
1. Turn Off the power to the TriPlus RSH system.
2. Move the Head manually in a position where it is easily accessible, preferably to an empty
slot of the Tool Station. Manually pull down the Z-axis to gain easy access to the lower
part.
3. Remove the front cover by loosening the screw at bottom of the cover. See
LED Color
LED Status
Off
Off
Solid On
Blinking fast
Blinking normal
Solid On
"Restoring Factory Default Network Settings"
6
Troubleshooting
Manually Removing a Tool from the Head
on
page
Figure
102.
TriPlus RSH Hardware Manual
77.
151