Maintaining a OC Injector Actuator

Maintaining a OC Injector Actuator
138
TriPlus RSH Hardware Manual
CAUTION Before any maintenance step is initiated, power down the TriPlus RSH.
Maintenance Step
Clean the outside of the instrument.
Use only a soft, lint-free cloth dampened
with mild soap and water.
Clean and dry the inside of the instrument if
necessary, including compartments, rolls and
rails. Use only a soft, lint-free cloth.
Clean and grease the rolls and the rails*.
*Recommended Grease
• Grease: AL Lub Kit: approximately 8 g
grease and brush
• Grease Temperature Range: -130 °C to
+150/+200 °C
Verify the position and condition of the
stopper and holding magnet of the
drawer(-s).
Condensation build-up is directly related to the temperature and relative humidity in the
ambient air (dew point). Long term tests have shown very little build-up of condensation in
an environment with relative humidity up to 60% and ambient temperatures of 22 °C ± 2 °C.
If condensation is observed, clean the compartment on a regular basis as needed.
CAUTION Before any maintenance step is initiated, power down the TriPlus RSH.
Maintenance Step
Verify the homing position of the OC
Actuator.
For further maintenance steps, refer to the
relevant GC's manual.
Interval
Weekly or as needed.
Daily or as needed.
After a year, if the system is not exposed to
acidic vapor.
If acids are used for the sample solution and
the Stack Cooler is exposed to the vapor,
cleaning should be done more frequently, as
required.
A positive stream of inert gas, such as
nitrogen, clean and dry air or helium, can
prevent corrosion, Condensation Build-up.
Monthly or as needed.
Interval
Weekly or as needed.
As required.