Placing The Triplus Rsh On The Supports - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

Placing the TriPlus RSH on the Supports

Thermo Scientific
2. Insert the provided fixing screw into each holes present on the support bar.
3. Mount each sampler support on the TRACE GC Ultra Pay attention that the support leg
is turned toward the back of the GC.
4. Guide the fixing screws into the corresponding fixing holes.
5. Tighten the fixing screws.
6. Place the two GC units beside the MS detector.
7. To mount the autosampler use the appropriate support and fixing it on the installation
holes provided on the top of the GC.
8. For further details please refer to the Operating Manual of the sampling system in use.
To place the TriPlus RSH onto the supports (vertical legs)
1. Loosen the two mounting clamps Torx screws of the vertical legs.
2. Place the TriPlus RSH base assembly on top and fit the legs into the groove in the X- axis.
CAUTION This operation must be performed by TWO persons who must stand each on
one side of the X-axis and put their hands underneath it.
3. Be sure that the mounting clamps fit completely into the grooves. Alternately tighten the
two Torx screws until the legs are firmly in place.
Figure 2.
Positioning and Fixing of the Mounting Clamps in the X-axis Grooves
4. Be sure that the mounting clamps fit completely into the grooves. Alternately tighten the
two Torx screws until the legs are firmly in place.
5. Double check if the leg mounting clamps are correctly hooked to the X-axis. See
2
Installation of the TriPlus RSH System
Placing the TriPlus RSH on the Supports
TriPlus RSH Hardware Manual
Figure
3.
19