3
TriPlus RSH Defining Object Positions
Teaching Standard Wash Station Position
Teaching Standard Wash Station Position
90
TriPlus RSH Hardware Manual
The teach position for the Standard Wash Station is the front vial (#5).
The Lower Needle Guide should be centered in these holes with the bottom of the lower
needle guide lightly touching the surface of the caps.
Figure 82. Teaching Position for Standard Wash Station and Fine-tuning of Vial #1
For the Standard Wash Station it is sufficient to teach position # 5 only. As long as the station
is mounted in square way, this approach is sufficient.
When in doubt, check each position individually. Select each position of the Standard Wash
Station, position #1 to position #5, and check the teach position. If a deviation is found, do
the necessary fine-tuning.
The teaching process for the Standard Wash Station with an inserted Waste Line Adapter does
not differ from the standard module. See
Figure 83. Teaching of Waste Positions with inserted Drainage Adapter
WARNING Observe safety measures if a waste line is installed in combination with a gas
chromatograph. Do not place the waste line above a heated zone.
Figure
83.
Thermo Scientific