Installing the Purge Gas Pressure Regulator

Thermo Scientific
WARNING Use an inert gas, such as helium or nitrogen, as purge gas.
When working with the connections for the pressure regulator, it is mandatory to always use
two flat wrenches as shown in
damaging or destroying the sealing gaskets.
Figure 20. Installation of the Purge Pressure Regulator (1)
To install the purge gas pressure regulator
1. Connect the OUT-side of the supplied pressure regulator at the back right side of the
TriPlus RSH. See
Figure
Figure 21. Installation of the Purge Pressure Regulator (2)
Figure
20. Using only one wrench increases the chances of
21.
2
Installation of the TriPlus RSH System

Installing the Purge Gas Pressure Regulator

TriPlus RSH Hardware Manual
31