Oc Injector Actuator Troubleshooting - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

OC Injector Actuator Troubleshooting

Thermo Scientific
Observation
Power Supply Problem
There are no 'operator-serviceable or replaceable parts inside the power supply. In case of
failure, contact your Thermo Fisher Scientific representative.
This section lists the most important points to overcome a faulty or problematic situation.
CAUTION Before starting any repair step, power down the TriPlus RSH.
Observation
The OC Actuator does not turn.
Needle Bending
Other items observed, such as bad
chromatographic results, may have other
sources which lead to the injector itself.
OC Injector Actuator Troubleshooting
Recommended Action
• Is the Temperature Controlled Drawer
connected to the same power module as
the TriPlus RSH?
If yes, use the second power module,
provided with the Temperature
Controlled Drawer.
• Is the power module connected, turned
on and the green LED ON (at outlet
side)?
If these points are verified but the problem
remains, ask your Thermo Fisher Scientific
representative for technical assistance.
Recommended Action
• Verify the electrical connections from
the OC Adapter to the OC Electronic
Box.
• Verify the connection from the OC
Electronic Box to the BUS connector.
• Verify any mechanical damage at the
OC Actuator.
• Check the teaching position of the OC
Injector.
• Consult the Manuals of the GC.
TriPlus RSH Hardware Manual
6
Troubleshooting
161