Teaching SPME Conditioning Station Position

Thermo Scientific
2. From the Options pull-up menu, select the item Setup | Modules | Tool Adapter | MHE
Station.
3. When the head has been positioned on the Adapter, which is placed in the MHE Adapter
Holder, the vent outlet should be placed in between the two pins.
4. Loosen the screw of the magnetic transport ring for aligning the magnetic transport ring
to the magnets of the MHE Adapter.
5. Define the X-, Y-, and Z- positions common to all objects.
The teach position is located at the front of the upper drawer.
The lower needle guide should be centered in the Lunette.
The SPME Conditioning Station teaching position is located as indicated in
Figure 92. Teaching Position for SPME Conditioning Station
To define the position of the SPME conditioning station
1. Select the access Extended User Level.
2. From the Options pull-up menu select the item Setup | Modules | SPME Conditioning
Station.
3. Select SPME Conditioning Station.
4. From the Options pull-up menu select the item Teach Module.
5. Connect the large vial transport ring to the lower needle guide.
3
TriPlus RSH Defining Object Positions

Teaching SPME Conditioning Station Position

Figure
TriPlus RSH Hardware Manual
92.
97