Teaching Large Volume Wash Station Position
Teaching Solvents Station Position
Thermo Scientific
The teaching position for the Large Volume Wash Station is the Waste.
The lower needle guide should be centered in these holes with the bottom of the Lower
Needle Guide slightly touching the surface of the Station assembly.
For the Large Volume Wash Station it is sufficient to teach the position Waste only. As long as
the module is mounted in square way, this approach is sufficient.
When in doubt, check each position individually. Select each position of the Large Volume
Wash Station, position #1 and position #2, and check the teach position. If a deviation is
found, do the necessary fine-tuning.
Figure 84. Teaching Position for Large Volume Wash Station and Fine-tuning.
WARNING Observe safety measures if a waste line is installed in combination with a gas
chromatograph. Do not position the waste line above a heated zone.
The teaching position for the Solvents Station is the third position.
The lower needle guide should be centered in these holes with the bottom of the Lower
Needle Guide slightly touching the surface of the reservoir bottle cap.
For the Solvent Station it is sufficient to teach position #3 only. As long as the module is
mounted in square way, this approach is sufficient.
When in doubt, check each position individually. Select each position of the Solvents Station,
position #1 and position #2, and check the teach position. If a deviation is found, do the
necessary fine-tuning.
3
TriPlus RSH Defining Object Positions
Teaching Large Volume Wash Station Position
TriPlus RSH Hardware Manual
91