Maintaining An Agitator; Maintaining A Vortexer - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

Maintaining an Agitator

Maintaining a Vortexer

Thermo Scientific
Maintenance Step
Check the flow of the solvent and check for
any blockage.
Check the cleanliness of the solvent frit in
the solvent reservoir bottle.
WARNING Before any maintenance step is initiated, power down the TriPlus RSH and
ensure that the temperature inside the unit is below 50 °C to avoid any burns.
Maintenance Step
Cleaning the outside of the Agitator:
Use only a soft lint-free cloth dampened with
mild soap and water.
Check opening and closing of the lid.
Move the lid manually back and forth to
ensure smooth movement.
As a standard procedure, the replacement of the agitator belt is recommended.
WARNING Before any maintenance step is initiated, power down the TriPlus RSH and
ensure that the temperature inside the unit is below 50 °C to avoid any burns.
Maintenance Step
Cleaning the outside of the Vortexer:
Use only a soft lint-free cloth dampened with
mild soap and water.
As a standard procedure, the replacement of the toothed motor belt is recommended.
Interval
Daily or as needed.
Daily or as needed.
Interval
Weekly or as needed.
Weekly or as needed.
Interval
Weekly or as needed.
TriPlus RSH Hardware Manual