Installation of the TriPlus RSH System

Installation of the TriPlus RSH System
Installation of the TriPlus RSH System
20
TriPlus RSH Hardware Manual
Figure 3.
Correct and Wrong Hooking of Mounting Clamps
This sections details the installation of the TriPlus RSH and its modules.
Note for Field Service Engineers — Discuss with the customer the positions of the TriPlus
RSH base and the TriPlus RSH modules hooked to the X-axis considering the GC
configuration. Open zones should be easily accessible by the TriPlus RSH Head but placing
a TriPlus RSH object above a heated zone could be dangerous, or at the very least influence
the stability of the analytical sample. Consider the space availability in the lab and to
adjacent instruments as well.
WARNING Do not place a TriPlus RSH module, which might contain flammable
solvents, directly above a heated zone. Special attention has to be paid to the Wash Station
and its waste line (if applicable), and Tray Holders with sample vials inserted.
Figure 4
shows an example of TriPlus RSH system with its modules installed.
Figure 4.
Example of TriPlus RSH System Modules
Correct
Wrong