Installing A Mhe Station - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

Installing a MHE Station

Thermo Scientific
2. Be sure that the clamps fit completely into the grooves. Alternately tighten the two Torx
screws until the two mounting clamps are firmly in place.
3. Double check if the Vortexer clamps are correctly hooked to the X-axis.
4. Connect the supplied control cable (daisy chain cable) from the connector on the back
side of the module to the Control Interface connector marked BUS. If BUS connector is
already occupied, connect the cable to another Active module with an open connector.
For details see the
Figure
Figure 32. Vortexer Electrical Connections
Before starting the installation process, determine the approximate position for the MHE
Adapter Holder.
The recommended position is directly next to the Agitator keeping about 40 mm distance.
If the Agitator is located at the far right side of the X-Axis, the MHE Adapter Holder should
be placed to the left side of the Agitator, or inside the path of travel.
Likewise, if the Agitator is located at the far left, the MHE Adapter Holder should be placed
to the right of the Agitator.
Note If a venting tube is installed, the MHE Adapter Holder must be installed as close as
possible to the Agitator (or any other position where the venting of the vial is performed)
to prevent the venting tube from catching on the hardware module.
To install the MHE station
1. Loosen the Torx screw on the mounting clamps located on the top side of the Fiber
Conditioning Station.
32.
2
Installation of the TriPlus RSH System

Installing a MHE Station

TriPlus RSH Hardware Manual
43