Installing The Fast Wash Station - Thermos TriPlus RSH Hardware Manual

Robotic sample handling
Table of Contents

Advertisement

Installing the Fast Wash Station

Thermo Scientific
The Fast Wash Station is shipped in a box separate from the system. The box also includes the
Solvent Reservoir Kit (two 1000 mL glass bottles, and a Tubing Kit).
CAUTION Do not place the station and specifically the waste line above a heated zone.
Tightly secure the waste line using a cable binder or similar item.
To install the fast wash station
1. Connect the solvent lines from the Solvent Reservoir Kit to the Inlet connectors of the
micro pump.
2. Loosen the two Torx screw on the mounting bracket for the module.
3. Hook the Fast Wash Station assembly to the X-axis.
4. Be sure that the clamp fits completely into the grooves. Alternately tighten the two Torx
screws until the mounting clamp is firmly in place.
5. Double check whether the Fast Wash Station clamp is correctly attached to the X-Axis.
Figure 26. Fast Wash Station
6. Connect the supplied control cable from the connector on the back side of the Fast Wash
Station to the Control Interface connector marked BUS. If BUS connector is already
occupied, connect the cable to another Active module with an open connector.