Maintenance
Cleaning of the LS/HS Syringe
Cleaning of the LS/HS Syringe
Cleaning the ITEX Trap
Maintaining a Fast Wash Station
134
TriPlus RSH Hardware Manual
The syringe cleaning is advisable every 100-200 samples, particularly if the sample analyzed
contains solids in suspension, or if poorly lubricant solvents are used, such as water and
methanol.
To clean the LS/HS syringe
1. Remove the syringe from the LS/HS tool.
2. Completely take out the syringe plunger and clean it gently using a soft cloth or a paper
tissue.
3. Reintroduce the syringe plunger and gently move it repeatedly along its whole travel
drawing an organic solvent (for example Hexane).
4. Reinstall the syringe in to the tool.
WARNING - BURN HAZARD Do not touch the surface of the Trap Heater during
operation to avoid burns. The surface temperature can reach potentially harmful levels
when operated at high desorbing or cleaning temperatures. The trap surface can reach
about 270 °C if operated at the maximum trap temperature of 350 °C.
To clean the ITEX trap
1. Clean a trap initially to eliminate all volatile contaminants. A practical check is to run a
blank injection, monitor the detector baseline to check for sufficient cleanness of the trap.
WARNING Before any maintenance step is initiated, power down the TriPlus RSH.
Maintenance Step
Cleaning the outside of the Fast Wash
Station: Use only a soft lint-free cloth
dampened with mild soap and water.
Check the integrity of the lines and
connections.
Interval
Weekly or as needed.
Daily or as needed.