Inserting a HS Syringe into the HS Tool

Thermo Scientific
To insert a HS syringe into the HS tool
1. Remove the retaining nut from the HS Tool, locate the O-ring, and put it over the syringe
glass barrel up to the flange. This is a protection for the glass flange. See
Figure 49. Preparing HS-Syringe for Installation
2. Insert the HS syringe into the HS Tool. Carefully insert the syringe body guiding the
needle tip into the upper and lower needle guide.
3. Press the syringe firmly into position, and tighten the syringe flange with the retaining
nut until a resistance is felt. The plunger is inserted in the syringe barrel.
Note It is not necessary to pay attention to the position of the syringe barrel side hole.
The heater design with the O-ring sealing allows a circular purge gas flow.
4. Check for correct installation. It is important that the needle tip is not visible outside of
the lower needle guide. See
Figure
50.
2
Installation of the TriPlus RSH System
Inserting a HS Syringe into the HS Tool
Figure
TriPlus RSH Hardware Manual
49.
55