5.13 After use

5.13.1 Switch device off completely
If the device is connected to a power supply, do not let it be turned off completely. To
completely disconnect the device from the power supply, turn the device to standby mode
and disconnect the device from the power supply. For optimal battery charging, however,
we recommend not to disconnect the device from the power supply.
To switch off the device completely in battery-operated mode, first switch to standby by
pressing the On/Off key for about 2 seconds until the blower switches off. Then press the
On/Off key again for at least 2 seconds until the device switches off completely and the
display goes out.
5.13.2 Mobile therapy data check
The therapy device has a memory card reader for MMC or SD cards which can be used to
save therapy data on a memory card. In discussion with the doctor supervising treatment,
this allows the patient's therapy data to be read out independently of the location of the
device, as the data can be transported on the memory card.
The following data are stored on the memory card:
– therapy pressure in hPa
– respiratory flow in l/min
– volume, leakage corrected, in ml
– current respiratory phase
– mean leakage flow in l/min
– current ratio Ti/T in %
– current respiratory frequency in 1/min
1. Switch the device to standby by keeping the On/Off
key depressed for approx. 2 seconds until the blower
switches off. The duration of the previous therapy
appears in the display. The device then switches to
standby.
2. Disconnect the patient/ventilator interface from the
patient circuit and the patient circuit from the device.
3. Clean the patient/ventilator interface, patient circuit
and device in accordance with the instructions for use.
For information on this, see the section entitled
"Hygiene treatment" on page 65.