4.4.2 Information relating to a separate exhalation system
Caution!
Always use an exhalation system. Used air containing carbon dioxide escapes from
the patient/ventilator interface (e.g. a mask) through the exhalation system.
Without an exhalation system, the CO
interface and the ventilation tube would rise to critical values and thus obstruct
breathing.
An exhalation system can either be integrated in the patient/ventilator interface, e.g. a
mask, or must be plugged in between the patient/ventilator interface and the patient
circuit in the form of an accessory.
If the patient/ventilator interface, e.g. the mask, does not have an exhalation system, a
separate exhalation system, e.g. Silentflow 2, must be used.
The exhalation system also allows a patient to breathe through his or her nose for a short
time, even if the device were to fail. In the case of full-face masks, breathing in the event
of a fault is through an emergency exhalation valve on the mask.
4.4.3 Connect separate exhalation system (only with leakage
ventilation)
The sleeve on the end of the patient circuit has a diameter of 19.5 mm and fits over a
22 mm standard tapered connector.
Patient/ventilator
interface
Exhalation system
Sleeve
Circuit
Follow the instructions for use for the exhalation system and for the patient/ventilator
interface.
5. Note that maximum flow rate and the accuracy of
dynamic pressure may deviate if you are not using
Weinmann tubes.
concentration in the patient/ventilator
2
To connect a separate exhalation system, proceed as
follows.
1. Plug the exhalation system into the sleeve of the circuit.
2. Connect the patient/ventilator interface to the
exhalation system.
EN
Set up device
41