If the therapy device is intended for use by several patients (e.g. in a hospital), a bacteria
filter must be used to prevent infections.
4.6.1 Leakage ventilation
WM 24148
1. Connect the VENTIclick humidifier directly to the therapy device.
2. Connect the bacteria filter to the outlet of the VENTIclick humidifier.
3. Connect the patient circuit to the bacteria filter.
4.6.2 Valve ventilation
WM 24476
If a humidifier and/or an oxygen sensor is to be connected as well, a different sequence
applies.
Combination with a humidifier
1. Connect the bacteria filter directly to the therapy device.
2. Connect the humidifier to the outlet of the bacteria filter.
3. Connect the patient circuit to the humidifier.
Combination with an oxygen sensor
1. Connect the oxygen sensor directly to the therapy device.
2. Connect the bacteria filter to the outlet of the oxygen sensor.
3. Connect the patient circuit to the bacteria filter.
In combination with leakage ventilation, use bacteria
filter WM 24148.
If the bacteria filter is used alone, it is connected directly
to the device outlet port and the ventilation tube is
plugged onto the bacteria filter.
If the VENTIclick humidifier is also to be connected,
proceed as follows.
In combination with valve ventilation, use bacteria
filter WM 24476.
If the bacteria filter is used alone, it is connected directly
to the device outlet port and the ventilation tube is
plugged onto the bacteria filter.
EN
Set up device
43