4. Set up device

Warning!
Device failure if incorrect patient circuits used!
If patient circuits with a diameter smaller than Ø 15 mm are used, the device may
overheat.
– Use only patient circuits with a diameter of Ø 15 mm or more.
– Note that total permitted resistance may be exceeded even in patient circuits
with a diameter of Ø 15 mm when these are combined with bacteria filters.
– Note on using Intersurgical
When using Intersurgical
single-patient circuits in which Intersurgical valves from the patient circuits mentio-
ned are used) the volume display on the device may deviate. The volume displayed
may deviate from that actually delivered by up to 105 ml.
– Use patient circuit WM 27181 from Weinmann to avoid this deviation in the
display.
The therapy device can be operated with both leakage and valve ventilation. The doctor
supervising treatment or the authorized specialist dealer converts the device to the variant
appropriate for the patient's form of treatment. The necessary settings on the device are
likewise made by the doctor supervising treatment.
You need only follow the relevant section depending on which patient circuit is being used.
Different functions and connection options for accessories are available depending on the
patient circuit. This is clearly indicated in the relevant sections.
4.1 Set up and connect the device
Set up the device on a flat surface, e.g. on a bedside cabinet or on the floor next to the
bed. Maintain a distance of at least 5 cm between the wall and the rear of the device, as
the air inlet is at the rear of the device. A distance of at least 5 cm should also be
maintained from the left-hand side of the device to allow the heat produced by the device
to escape.
®
patient circuits
®
patient circuits ref. 5183064 and ref. 5083 (and other
EN
Set up device
37