Disposal - Weinmann VENTIlogic LS Instructions For Use Manual

Ventilogic ls - single patient circuit with patient valve; ventilogic plus - single patient circuit with patient valve
Hide thumbs Also See for VENTIlogic LS:
Table of Contents

Advertisement

Available languages

Available languages

or
• after 2 years (see service label on the rear of the device)
depending on which comes first.
The safety check and servicing include:
check for completeness
visual inspection for mechanical damage
filter change
cleaning the device
replacement of any defective parts
complete check of device functions and pressure displays
battery change
change of internal battery
final check in accordance with test instruction WM 27104
testing the power supply failure alarm

9.6 Disposal

9.6.1 Device
9.6.2 Internal battery
9.6.3 Oxygen sensor
Do not dispose of the device with domestic waste. To
dispose of the device properly, contact a licensed,
certified electronic scrap disposal merchant. This address
is available from your Environment Officer or from your
local authority.
Do not dispose of the internal battery with domestic
waste. To dispose of the battery properly, contact a
licensed, certified electronic scrap disposal merchant.
This address is available from your Environment Officer or
from your local authority.
Do not dispose of the oxygen sensor with domestic
waste, as it contains electrolyte and lead. It should be
disposed of in accordance with applicable national or
regional regulations.