Available languages

7.2.3 Check flow measurement and flow sensors/pressure sensors
(leakage ventilation with leakage adapter for standard
tapered connector)
7.2.4 Check flow measurement and flow sensors/pressure sensors
(valve ventilation)
Function check using a single patient circuit with patient valve
6. If the function check is completed successfully, the
message Function check ok! appears.
If the function check is not successful, the message
Device system implausible appears. In this case
refer to section "8.1 Faults" on page 79.
7. Press the menu key to return to the standard display.
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. Perform the subsequent steps as descibed in
"7.2.2 Check flow measurement and flow sensors/
pressure sensors (leakage ventilation)" from step 2.
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. The subsequent steps are described in "Check flow
measurement and flow sensors/pressure sensors
(leakage ventilation)" starting from step 2.
EN
Function check
73