Clean Leakage Ventilation - Weinmann VENTIlogic LS Instructions For Use Manual

Ventilogic ls - single patient circuit with patient valve; ventilogic plus - single patient circuit with patient valve
Hide thumbs Also See for VENTIlogic LS:
Table of Contents

Advertisement

Available languages

Available languages

Interval
Annually
Follow the relevant instructions for use on the hygiene treatment for the patient/ventilator
interface.
6.1.2 Valve ventilation
Interval
Daily
Weekly
Every 1000 operating
hours
Every 6 months
The patient circuits for single and double patient circuits with patient valve (double patient
circuit only on VENTIlogic LS) are disposables and cannot be subjected to a hygiene
treatment. Follow the instructions for use for the patient circuit in question.
Follow the relevant instructions for use for the hygiene treatment of the patient/ventilator
interface.

6.2 Clean leakage ventilation

6.2.1 Clean patient circuit
1. Pull the patient circuit off the device and the exhalation system.
66
EN
Hygiene treatment
– Change patient circuit.
– Clean the humidifier in accordance with the relevant
instructions for use.
– Change bacteria filter WM 24476.
– Clean coarse dust filter.
– Clean fan filter.
– Change fine filter (filter change indicator
– Change coarse dust filter, earlier if dirty or worn.
– Change fan filter.
2. Pull out the one end of the pressure measuring tube
(shake a little if necessary) and seal it with the sealing
plug supplied. At the other end, seal the small opening
of the adapter using the second sealing plug so that
no water can penetrate.
3. Clean the creased tube with a little detergent in hot
water and make sure no residues are left behind. Flush
the inside of the tube through thoroughly in the
process.
Activity
Activity
), earlier if dirty.