Weinmann VENTIlogic LS Instructions For Use Manual page 14

Ventilogic ls - single patient circuit with patient valve; ventilogic plus - single patient circuit with patient valve
Hide thumbs Also See for VENTIlogic LS:
Table of Contents

Advertisement

Available languages

Available languages

VENTIlogic LS:
3
Connection for patient's exhaled air with double patient circuit with patient valve.
VENTIlogic plus:
Opening is not used with VENTIlogic plus.
4
Jack: electrical connection for VENTIclick respiratory air humidifier (only leakage
ventilation), max. current consumption at 40 V: 600 mA
5
Jack: electrical connection for oxygen sensor; max. 100 mV DC
6
Connection: pressure measuring tube (marked blue). Therapy pressure 0 - 50 hPa
P
(only for patient circuit with patient valve)
7
Connection: control tube for patient valve 0 - 50 hPa (only patient circuit with
patient valve)
8
Device outlet port: outlet for exhaled air at 0 - 45 hPa with patient circuits with
patient valve, 0 - 35 hPa with leakage ventilation
9
Device outlet port: only patient circuits with a diameter of Ø 15 mm - 22 mm are
permitted.
10
Connection for optional attachments, e.g. Analog box D/A;
max. current delivery at 5 V: 50 mA
11
Connection for specialist staff to set therapy parameters using
WEINMANNsupport; max. current delivery at 12 V: 50 mA
12
Slot for MMC or SD card
13
Connection for power supply-independent operation with VENTIpower
~
14
Connector for power supply input 115/230 V AC; 50/60 Hz
15
Connection for remote alarm: connection for nurse call system and VENTIremote
alarm remote alarm case. Breaking capacity: 60 V DC/2 A; 42 V AC/2 A
14
EN
Overview
Front
Right-hand side
Rear