3.1.3 Mobile operation
Warning!
•When operated on a wheelchair, this system requires a certificate of conformity.
To obtain this, consult a specialist dealer authorized by the manufacturer,
Weinmann; this dealer will also perform the attachment.
Caution!
• The internal battery is not intended for mobile operation. Always use
VENTIpower for a mobile power supply.
• Protect the therapy device in the carrying bag from direct sunlight and rain,
using a sunshade or rain cover for example. The carrying bag itself provides only
brief protection from sunlight and rain.
• During mobile operation there may be problems with the trigger. This can result
in inadequate ventilation. Have your physician adjust the trigger settings or
select a monitored ventilation mode.
Note:
• When planning your time, take account of the fact that battery running time is
considerably reduced at low outdoor temperatures.
3.1.4 Oxygen supply
Warning!
•If oxygen is being supplied to the respiratory flow, smoking and naked flames
are forbidden. Risk of fire. The oxygen can accumulate in clothing, bed linen or
hair. It can only be removed by thorough ventilation.
• It is essential to follow the safety instructions in the instructions for your oxygen
system.
• Too high or too low an oxygen supply can be toxic and lead to severe
complications. We therefore recommend monitoring the oxygen supply with an
oxygen sensor. This oxygen sensor can replace neither blood gas analysis nor
direct FiO
• Risk of fire. Always shut off the oxygen supply first at the end of therapy. Then
leave the therapy device to run for a short time before turning it off, otherwise
some residual oxygen will remain in the device. In the event of a fault, this could
lead to a risk of fire.
measurement.
2