Flow rates for the Xevo G2 QTof system

The Xevo G2 QTof system can run at high flow rates. To optimize desolvation,
and thus sensitivity, run the system at appropriate gas flows and desolvation
temperatures.
Flow rate versus temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020
0.020 to 0.100
0.101 to 0.300
0.301 to 0.500
>0.500
Preparing the IntelliStart Fluidics system
For additional information, see
page
B-15.
Warning:
exposure to toxic substances, never place containers on top of the
instrument or on its front covers.
Installing the reservoir bottles
Use standard reservoir bottles (30-mL) for instrument setup and calibration.
Use the Low-volume Adaptor Kit (included) to infuse smaller volumes. The
low-volume vials have a volume of 1.5 mL.
Required materials
Chemical-resistant, powder-free gloves
2-4
Preparing the Mass Spectrometer for Operation
Source
temperature (°C)
100
120
120
150
150
"Connecting the liquid waste line" on
To avoid injuries from broken glass, falling objects, or
Desolvation
temperature (°C)
200
350
450
500
0
Desolvation gas
flow (L/h)
800
800
800
1000
1200