Installing The New Capillary - Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

7.
Unscrew the finger-tight PEEK union from the UNF coupling.
8.
Remove the ferrule from the capillary.
9.
Remove the capillary from the UNF coupling.
10. Dispose of the capillary and ferrule in accordance with local
environmental regulations.

Installing the new capillary

Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
7-mm wrench
Combined 2.5-mm Allen wrench and cone extraction tool
Red PEEK tubing
LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Sharp knife or PEEK tubing cutter
Safety goggles
5-78
Maintenance Procedures
Ferrule
The capillary and ferrule can be contaminated
Warning:
with biohazardous and/or toxic materials. Dispose of them
according to local environmental regulations.