Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual page 215

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

2.
Place the pump on the PTFE drip tray.
3.
Using an NW25 center ring, attach the flanged end of the vacuum hose
to the roughing pump inlet flange, and then secure the connection with
an NW25 clamp, tightening it with the 7-mm nut driver.
4.
Connect the free end of the vacuum hose to the vacuum port on the mass
spectrometer's rear panel.
Caution:
exhaust tubing squarely (that is, perpendicular to its horizontal
axis).
5.
Using an NW25 center ring, attach the 12.7-mm clear PVC exhaust
tubing to the roughing pump oil exhaust filter port NW25 nozzle fitting,
and then secure the connection with an NW25 clamp, tightening it with
the 7-mm nut driver.
6.
Secure the exhaust tubing with a hose clamp.
Caution:
one for nitrogen, the other for the roughing pump. Oil mist can
seriously damage the instrument if the nitrogen exhaust line
connects with the roughing pump exhaust line. Your warranty
does not cover damage caused by routing exhaust lines incorrectly.
See also
"Connecting the nitrogen exhaust line" on page
7.
Route the open end of the exhaust tubing to a suitable exhaust vent. For
further information, see the Waters Xevo G2 QTof MS Site Preparation
Guide (part number 715001948).
Caution:
operate the pump with the oil level outside the range shown by the
oil level sight glass minimum and maximum oil level indicators.
8.
Inspect the oil level in the pump (see
pump oil level" on page
roughing pump" on page
9.
Make the electrical connections to the roughing pump (see
To avoid gas leaks, use the sharp knife to cut the PVC
The instrument requires two separate exhaust systems:
To ensure correct operation of the roughing pump, do not
5-20, and, if necessary,
5-21).
Connecting the Varian oil-filled roughing pump
"Inspecting the Varian roughing
"Adding oil to the Varian
B-12.
page
B-6).
B-5