Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual page 193

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

Required materials
Chemical-resistant, powder-free gloves
The Xevo G2 QTof Fluidics Tubing and Fitting Kit
This kit contains components for both the sample and lock-spray
Tip:
system's plumbing.
Needle-nose pliers
To plumb the analyte system:
Warning:
contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while working on the
system.
Finger tighten all PEEK fittings.
Requirement:
1.
Open the access doors to the IntelliStart Fluidics system.
Caution:
end. Ensure that you tighten the fittings appropriately.
2.
Using the stainless steel tubing, connect the sample pump to port 7 of
the sample selector valve.
A
B
C
Lock-spray pump
The IntelliStart Fluidics tubing and fittings can be
The steel tubing requires high-pressure seals at each
A
B
C
C
B
A
Wa ters
B
C
A
Replacing the IntelliStart Fluidics tubing 5-117
Sample selector valve