To remove the source enclosure:
1.
Prepare the instrument for working on the source (see
2.
Remove the probe from the source:
If you are removing an ESI probe, see
If you are removing an APCI probe, see
3.
Slide open the instrument's source interface door (see the figure on
page
1-4).
4.
Disconnect the probe adjuster and options cables from the instrument's
connectors.
5.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
5-8
Maintenance Procedures
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The probe and source can be hot. To avoid burn injuries,
Warning:
take great care while working with these components.
The corona pin tip is sharp. To avoid puncture wounds,
Warning:
take great care while working with the source enclosure open if a
corona pin is fitted.
If you are removing a NanoLockSpray source enclosure,
Caution:
to avoid damage to the sample inlet, you must move the sprayer
platform out of the source enclosure before you open the source
enclosure (see
page
page
page
4-6).
page
5-7).
3-6.
3-17.