External connections 1:
Pin
Function
1
Event In 1+, digital signal,
optimum +3.3V max +5V
2
Event In 1-, digital ground, 0V
3
Not used
4
Event In 2+, digital signal,
optimum +3.3V max +5V
5
Event In 2-, digital ground, 0V
6
Not used
7
Not used
8
CE Interlock Out, common
9
CE Interlock Out, normally closed (N/C)
10
CE Interlock Out, normally open (N/O)
External connections 2:
Pin
Function
1
Analog Out +, instrument-driven electrical
output
2
Analog Out -, ground
3
Gas Fail Interlock, common
4
Gas Fail Interlock, normally closed (N/C)
5
Gas Fail Interlock normally open (N/O)
6
Event Out 1 +
7
Event Out 1 -
8
Not used
9
Event Out 2 +
10
Event Out 2 -
B-22
External Connections
Rating
+5 V
0 V
+5 V
0 V
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA
Rating
None
None
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA
+30 V d.c., 100 mA