Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual page 153

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

3.
Remove the end cover and gasket.
Nebulizer adjuster knob
4.
Unscrew and remove the nebulizer adjuster knob.
5.
Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
PEEK union/UNF coupling assembly
Locknut
6.
Use the 7-mm wrench to loosen the locknut.
Gasket
End cover
Capillary
Replacing the APCI probe sample capillary
5-77